Cytek Aurora CS (S0366SP)

Cytek Aurora CS

Cytek Aurora CS (S0366SP)|Configuration|Fluorescent Guide (Cytek)

  • 5 lasers: 20mW Ultraviolet (355nm), 100mW Violet (405nm), 50mW Blue (488nm), 50mW Yellow-Green (561nm), 80mW Red (640nm) spatially-separated laser sources
  • 64 fluorescent detectors (16UV-16V-14B-10YG-8R)
  • Semiconductor array for each laser enabling more efficient spectrum capture in the 365nm-829nm range. No filter changes required for any fluorophores excited by the 355nm, 405nm, 488nm, 561nm and 640nm lasers.
  • SpectroFlo® software can be run in full spectrum mode to take advantage of the full spectrum of each fluorophore or run in conventional compensation mode to assign each fluorophore to a specific detector.
  • Equipped with Enhanced Small Particle (ESP) Detection Option capable of resolving particles as small as 70nm
  • Pulse Area and Height for all parameters; Pulse Width of scatter parameters and one fluorescent parameter per laser
  • Variable speed sorter with 70µm, 85µm, 100µm and 130µm nozzles
  • Variable sample and sorted cell temperature control (4°C, 10°C, 15°C, 20°C, 25°C, 30°C, 37°C)
  • Up to 2-way sorting (15mL conical tubes) or 6-way sorting (5mL round-bottom or 1.5mL Eppendorf tubes)
  • Plate sorting (96-well or 384-well plates) with single cell, index sorting capability
  • In-line end of sample detector to detect bubbles before they reach the flow cell
  • BSL-2E compliant sorter installed inside a Baker Bio-Protect III biohazard containment cabinet with an integrated Aerosol Management System (AMS)
  • Accessible to the MRF BSL-2 Enhanced Facility (2-126 MRF) through pass-through window to facilitate easy sorting access when working with HIV, SIV or other BSL-2E pathogens

Location: 2-128A Microbiology Research Facility (MRF)